首页 / 产品 / ELISA试剂盒 / 人ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human RIPK3 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 98 | ||||
Range % | 94-103 | |||||
1:2 | Average % | 95 | ||||
Range % | 90-102 | |||||
1:4 | Average % | 89 | ||||
Range % | 84-94 | |||||
1:8 | Average % | 90 | ||||
Range % | 85-96 |
The recovery of human RIPK3 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 93 | 86-98 | ||||
EDTA plasma (n=4) | 87 | 80-95 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
10 | 2.512 | 2.634 | 2.573 | 2.408 | ||
5 | 1.818 | 1.767 | 1.793 | 1.627 | ||
2.5 | 1.305 | 1.328 | 1.317 | 1.151 | ||
1.25 | 0.806 | 0.829 | 0.817 | 0.652 | ||
0.625 | 0.524 | 0.521 | 0.522 | 0.357 | ||
0.312 | 0.347 | 0.353 | 0.350 | 0.185 | ||
0.156 | 0.263 | 0.254 | 0.258 | 0.093 | ||
0 | 0.166 | 0.165 | 0.165 |
A micro ELISA plate --- The 96-well plate has been pre-coated with an anti-human RIPK3 antibody. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard --- Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
Biotin-labeled RIPK3 antibody (100 x concentrate) 1 x 120 μl ---Act as the detection antibody.
HRP-avidin (100 x concentrate) 1 x 120 μl --- Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
Biotin-antibody Diluent 1 x 15 ml ---Dilute the high concentration Biotin-antibody to an appropriate working solution.
HRP-avidin Diluent 1 x 15 ml ---Dilute the high concentration HRP-avidin solution to an appropriate solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator that can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×