Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human L-Carnitine in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:200 | Average % | 88 | ||||
Range % | 80-96 | |||||
1:400 | Average % | 97 | ||||
Range % | 91-105 | |||||
1:800 | Average % | 99 | ||||
Range % | 92-110 | |||||
1:1600 | Average % | 100 | ||||
Range % | 90-110 |
The recovery of human L-Carnitine spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 93 | 82-110 | ||||
EDTA plasma (n=4) | 97 | 90-110 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
20 | 2.749 | 2.639 | 2.694 | 2.543 | ||
10 | 2.010 | 1.976 | 1.993 | 1.842 | ||
5 | 1.303 | 1.294 | 1.299 | 1.148 | ||
2.5 | 0.889 | 0.839 | 0.864 | 0.713 | ||
1.25 | 0.561 | 0.524 | 0.543 | 0.392 | ||
0.625 | 0.371 | 0.359 | 0.365 | 0.214 | ||
0.312 | 0.256 | 0.239 | 0.248 | 0.097 | ||
0 | 0.155 | 0.147 | 0.151 | |||
×