Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-5 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 82 | ||||
Range % | 80-92 | |||||
1:2 | Average % | 89 | ||||
Range % | 85-100 | |||||
1:4 | Average % | 92 | ||||
Range % | 90-105 | |||||
1:8 | Average % | 90 | ||||
Range % | 86-98 |
The recovery of mouse IL-5 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 88 | 82-93 | ||||
EDTA plasma (n=4) | 87 | 80-94 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
1000 | 2.346 | 2.434 | 2.390 | 2.244 | ||
500 | 1.725 | 1.767 | 1.746 | 1.600 | ||
250 | 1.285 | 1.328 | 1.307 | 1.161 | ||
125 | 0.910 | 0.929 | 0.920 | 0.774 | ||
62.5 | 0.553 | 0.521 | 0.537 | 0.391 | ||
31.2 | 0.325 | 0.343 | 0.334 | 0.188 | ||
15.6 | 0.216 | 0.204 | 0.210 | 0.064 | ||
0 | 0.147 | 0.145 | 0.146 |
×