Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of human N-Cad in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 88 | ||||
Range % | 80-100 | |||||
1:2 | Average % | 97 | ||||
Range % | 91-104 | |||||
1:4 | Average % | 97 | ||||
Range % | 92-108 | |||||
1:8 | Average % | 94 | ||||
Range % | 85-98 |
The recovery of human N-Cad spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 95 | 89-105 | ||||
EDTA plasma (n=4) | 97 | 90-110 | ||||
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
400 | 2.152 | 2.174 | 2.163 | 2.005 | ||
200 | 1.832 | 1.824 | 1.828 | 1.670 | ||
100 | 1.220 | 1.251 | 1.236 | 1.078 | ||
50 | 0.926 | 0.974 | 0.950 | 0.792 | ||
25 | 0.574 | 0.543 | 0.559 | 0.401 | ||
12.5 | 0.406 | 0.426 | 0.416 | 0.258 | ||
6.25 | 0.335 | 0.312 | 0.324 | 0.166 | ||
0 | 0.152 | 0.164 | 0.158 | |||
×