Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse TNF-α in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 88 | ||||
Range % | 85-92 | |||||
1:2 | Average % | 94 | ||||
Range % | 91-97 | |||||
1:4 | Average % | 99 | ||||
Range % | 92-110 | |||||
1:8 | Average % | 96 | ||||
Range % | 89-99 |
The recovery of mouse TNF-α spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 95 | 90-102 | ||||
EDTA plasma (n=4) | 97 | 94-105 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
4000 | 2.727 | 2.634 | 2.680 | 2.509 | ||
2000 | 2.027 | 1.967 | 1.997 | 1.826 | ||
1000 | 1.236 | 1.158 | 1.197 | 1.026 | ||
500 | 0.742 | 0.729 | 0.735 | 0.564 | ||
250 | 0.428 | 0.421 | 0.424 | 0.253 | ||
125 | 0.366 | 0.353 | 0.359 | 0.188 | ||
62.5 | 0.268 | 0.274 | 0.271 | 0.100 | ||
0 | 0.167 | 0.175 | 0.171 |
×