Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human IL-22 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 103 | ||||
Range % | 99-109 | |||||
1:2 | Average % | 109 | ||||
Range % | 105-115 | |||||
1:4 | Average % | 98 | ||||
Range % | 91-105 | |||||
1:8 | Average % | 93 | ||||
Range % | 86-100 |
The recovery of human IL-22 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 107 | 103-111 | ||||
EDTA plasma (n=4) | 99 | 92-106 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
5000 | 2.994 | 3.064 | 3.029 | 2.921 | ||
2500 | 2.338 | 2.307 | 2.322 | 2.215 | ||
1250 | 1.400 | 1.358 | 1.379 | 1.272 | ||
625 | 0.719 | 0.729 | 0.724 | 0.617 | ||
312 | 0.371 | 0.391 | 0.381 | 0.274 | ||
156 | 0.271 | 0.263 | 0.267 | 0.160 | ||
78 | 0.196 | 0.184 | 0.190 | 0.083 | ||
0 | 0.110 | 0.105 | 0.107 |
A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-human IL-22 antibody. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial Biotin-labeled IL-22 antibody (100 x concentrate) (120 μl/bottle) ---Act as the detection antibody.
One vial HRP-avidin (100 x concentrate) (120 μl/bottle) ---Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
One vial Biotin-antibody Diluent (15 ml/bottle) ---Dilute the Biotin-antibody.
One vial HRP-avidin Diluent (15 ml/bottle) ---Dilute the HRP-avidin solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×