首页 / 产品 / ELISA试剂盒 / 大鼠ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of rat IL-1α in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 96 | ||||
Range % | 92-102 | |||||
1:2 | Average % | 98 | ||||
Range % | 91-104 | |||||
1:4 | Average % | 101 | ||||
Range % | 98-107 | |||||
1:8 | Average % | 89 | ||||
Range % | 84-96 |
The recovery of rat IL-1α spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 98 | 92-106 | ||||
EDTA plasma (n=4) | 95 | 89-99 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
5000 | 2.657 | 2.556 | 2.607 | 2.425 | ||
2500 | 2.123 | 2.022 | 2.073 | 1.891 | ||
1250 | 1.468 | 1.447 | 1.458 | 1.276 | ||
625 | 0.927 | 0.916 | 0.922 | 0.740 | ||
312 | 0.568 | 0.579 | 0.574 | 0.392 | ||
156 | 0.415 | 0.404 | 0.410 | 0.228 | ||
78 | 0.312 | 0.300 | 0.306 | 0.124 | ||
0 | 0.180 | 0.183 | 0.182 |
×