首页 / 产品 / ELISA试剂盒 / 小鼠ELISA试剂盒
| Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<10% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of mouse COL18A1/ES in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:1 | Average % | 92 | ||||
| Range % | 89-95 | |||||
| 1:2 | Average % | 103 | ||||
| Range % | 99-105 | |||||
| 1:4 | Average % | 92 | ||||
| Range % | 87-96 | |||||
| 1:8 | Average % | 99 | ||||
| Range % | 94-101 | |||||
| The recovery of mouse COL18A1/ES spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 95 | 91-98 | ||||
| EDTA plasma (n=4) | 95 | 91-98 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
| ng/ml | OD1 | OD2 | Average | Corrected | ||
| 10 | 2.234 | 2.213 | 2.224 | 2.067 | ||
| 5 | 1.732 | 1.694 | 1.713 | 1.556 | ||
| 2.5 | 1.091 | 1.076 | 1.084 | 0.927 | ||
| 1.25 | 0.551 | 0.537 | 0.544 | 0.387 | ||
| 0.625 | 0.366 | 0.386 | 0.376 | 0.219 | ||
| 0.312 | 0.292 | 0.281 | 0.287 | 0.130 | ||
| 0.156 | 0.248 | 0.228 | 0.238 | 0.081 | ||
| 0 | 0.159 | 0.155 | 0.157 | |||
×