Intra-assay Precision (Precision within an assay): CV%<12% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<15% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of rat ANG-Ⅱ in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 84 | ||||
Range % | 80-92 | |||||
1:2 | Average % | 87 | ||||
Range % | 81-95 | |||||
1:4 | Average % | 90 | ||||
Range % | 92-97 | |||||
1:8 | Average % | 93 | ||||
Range % | 86-98 |
The recovery of rat ANG-Ⅱ spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 92 | 86-98 | ||||
EDTA plasma (n=4) | 87 | 82-95 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | |||
1000 | 0.158 | 0.160 | 0.159 | |||
500 | 0.185 | 0.189 | 0.187 | |||
250 | 0.237 | 0.232 | 0.235 | |||
125 | 0.364 | 0.340 | 0.352 | |||
62.5 | 0.595 | 0.568 | 0.581 | |||
31.25 | 0.958 | 0.902 | 0.930 | |||
15.6 | 1.355 | 1.298 | 1.326 | |||
0 | 2.287 | 2.126 | 2.206 |
×