Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human versican/PG-M/PG-350 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:20 | Average % | 90 | ||||
Range % | 85-95 | |||||
1:40 | Average % | 97 | ||||
Range % | 91-105 | |||||
1:80 | Average % | 92 | ||||
Range % | 88-97 | |||||
1:160 | Average % | 93 | ||||
Range % | 86-98 |
The recovery of human versican/PG-M/PG-350 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 95 | 89-98 | ||||
EDTA plasma (n=4) | 90 | 84-95 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
30 | 2.803 | 2.734 | 2.768 | 2.595 | ||
10 | 2.365 | 2.367 | 2.366 | 2.193 | ||
3.33 | 1.805 | 1.828 | 1.816 | 1.644 | ||
1.11 | 1.114 | 1.129 | 1.121 | 0.948 | ||
0.37 | 0.673 | 0.651 | 0.662 | 0.489 | ||
0.12 | 0.384 | 0.363 | 0.373 | 0.200 | ||
0.04 | 0.265 | 0.274 | 0.269 | 0.096 | ||
0 | 0.171 | 0.175 | 0.173 |
×