Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. | ||||||
Intra-Assay PrecisionInter-Assay PrecisionSample123123n202020202020Mean(pg/ml)18.59618.52918.51218.42618.54618.495SD0.0410.0460.0440.0480.0560.054CV(%)3.0103.3873.2423.5514.1213.982 | ||||||
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-1α in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
SampleSerum(n=4)1:1Average %99Range %93-1051:2Average %92Range %87-991:4Average %95Range %90-991:8Average %93Range %88-96 | ||||||
The recovery of mouse IL-1α spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample TypeAverage % RecoveryRangeSerum (n=5)9390-98 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/mlOD1OD2AverageCorrected1502.9012.8872.8942.72752.4612.3452.4032.22937.51.9711.8281.9001.75518.751.3621.2781.321.1469.380.7170.6890.7030.5294.690.4580.4470.4530.2792.350.2980.2810.290.11600.1750.1730.174 |
A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-mouse IL-1α antibody. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial Biotin-labeled IL-1α antibody (100 x concentrate) (120 μl/bottle) ---Act as the detection antibody.
One vial HRP-avidin (100 x concentrate) (120 μl/bottle) ---Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
One vial Biotin-antibody Diluent (15 ml/bottle) ---Dilute the Biotin-antibody.
One vial HRP-avidin Diluent (15 ml/bottle) ---Dilute the HRP-avidin solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×