首页 / 产品 / ELISA试剂盒 / 大鼠ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess.
Inter-assay Precision (Precision between assays): CV%<10%
Three samples of known concentration were tested in twenty assays to assess.
To assess the linearity of the assay, samples were spiked with high concentrations of rat IGF-1 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
A micro ELISA plate --The 96-well plate has been pre-coated with an anti-rat IGF-1 antibody. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard --Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial Biotin-labeled IGF-1 antibody (100 x concentrate) (120 μl/bottle) --Act as the detection antibody.
One vial HRP-avidin (100 x concentrate) (120 μl/bottle) --Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
One vial Biotin-antibody Diluent (15 ml/bottle) --Dilute the high concentration Biotin-antibody to an appropriate working solution.
One vial HRP-avidin Diluent (15 ml/bottle) --Dilute the high concentration HRP-avidin solution to an appropriate solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×